GfG O2 Replacement Sensor 0 - 25% vol. 1990716
We proudly serve diverse industries including construction, oil and gas, confined space operations, manufacturing, and more. Contact our California-based safety experts or give them a call for personalized guidance on your safety needs.
Description
Description
The GfG O2 Replacement Sensor (1990716) is a drop-in electrochemical oxygen sensor for the GfG G222E Micro 5, G888, and G999 instrument families. It reads 0–25% O2 by volume at 0.01% resolution, which is the standard window for confined-space oxygen monitoring where OSHA defines oxygen-deficient as below 19.5% and oxygen-enriched as above 23.5%. Replace at or before the 2-year warranty interval to keep life-safety alarms firing accurately.
Key Specifications:
- Target gas: Oxygen (O2), 0–25% by volume
- Resolution: 0.01% O2
- Sensor type: Electrochemical
- Compatible with: GfG G222E Micro 5, G888, G999
- Warranty: 2 years
Compatible GfG instruments:
- GfG G222E Micro 5 Single Gas O2 Detector
- GfG G888 portable multi-gas monitor
- GfG G999 portable multi-gas monitor
Part Number:
1990716 — O2 Replacement Sensor, 2-year warranty
Frequently Asked Questions
How long does an electrochemical O2 sensor last?
Galvanic electrochemical O2 sensors typically last 2–3 years under normal operating conditions, which matches the 2-year warranty on this unit. Service life is shortened by sustained exposure to oxygen-enriched atmospheres, corrosive gases, or temperatures outside the operating range. Replace at or before the rated interval regardless of apparent performance.
Does the sensor need calibration after installation?
Yes. After installing a fresh O2 sensor, let the instrument stabilize per the GfG manual, then calibrate zero and span. A simple fresh-air zero (20.9% O2) is the minimum; a full span calibration using a known O2 standard is the correct full procedure. A bump test alone is not a substitute for calibration after a sensor swap.
What is OSHA's definition of oxygen-deficient and oxygen-enriched?
OSHA 29 CFR 1910.146 classifies atmospheres below 19.5% O2 as oxygen-deficient and above 23.5% O2 as oxygen-enriched, both outside the acceptable range for normal work in a permit-required confined space. The regulator relies on the monitor's O2 sensor reading accurately, which is why sensor replacement at or before the rated interval is non-negotiable.
Can PK Safety service my GfG instrument?
PK Safety stocks GfG replacement sensors and accessories and can walk you through the swap and calibration procedure. For factory-level repair or warranty service on GfG instruments, call us at 800.829.9580 and we'll route the instrument appropriately.
Specifications
Specifications
-
MPN
-
Product Weight (lbs.)
-
Country of Origin
Reviews
Reviews
Keep your team protected & compliant!
- Expert guidance tailored to your specific needs
- Access responsive customer support when you need it
- Enjoy fast shipping directly to your location
- Benefit from extensive industry knowledge—75+ years of experience
- In-house gas detection calibration & repair services available